Crest Electronics CDVS-7000 Manuel d'utilisateur Page 156

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 341
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 155
User Manual
142 i Series / iX Series
FREQUENCY TRANSIENTS (Group 1 only)
Seq. No
Frequency
Time
1
400
5 Minute
2
440
150msec
3
400
5Sec.
4
350
150msec
5
400
5Sec.
Table 9-5: Normal Frequency Transient Sequence
This test applies to Group 1 only. At 115 voltage, change the frequency per sequence listed in
Table 9-5. The test will cycle 5 times starting from sequence 2.
FREQUENCY VARIATION (Group 2 and 3 only)
Seq. No
Initial Frequency
Slew rate
Final Frequency
Group2
Group3
Hz/Sec
Group2
Group3
1
360
360
100
650
800
2
650
800
100
360
360
3
360
360
Pause 5 sec
360
360
Table 9-6: Normal Frequency Variation Sequence
This test will apply to Group 2 and 3 only. At 115 voltage, the frequency is set to 360Hz for 5
minutes. The frequency is slowed per sequence listed in Table 9-6. The test will cycle 3 times.
The frequency will return to nominal after the test is completed.
9.1.1.5.2 EMERGENCY TEST
From the DO160 MENU scroll to the EMERGENCY AC entry using the up and down cursor
keys. Press the ENTER key to select the EMERGENCY screens. The screen will appear as
shown in Figure 9-8.
Figure 9-8: Emergency Screen
The EMERGENCY SCREEN has the following tests:
1
VOLT FREQ MIN
2
VOLT FREQ MAX
3
VOLT UNBALANCE
The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.
Vue de la page 155
1 2 ... 151 152 153 154 155 156 157 158 159 160 161 ... 340 341

Commentaires sur ces manuels

Pas de commentaire